基于CPLD高精度恒流源技术的呼吸体积描记系统的设计
DOI:
作者:
作者单位:

(1.解放军第305医院 医学工程科, 北京 100017;2.解放军第305医院 呼吸内科,北京 100017; ;3.南方医科大学 生物医学工程学院, 广州 510515)

作者简介:

姬 军(1970),男,北京人,博士,副主任技师,主要从事生物医学信号检测与处理等方向的研究。

通讯作者:

中图分类号:

R197.39

基金项目:

首都医学发展基金资助项目(2007-3038)。


Design of a Respiratory Inductive Plethysmography System Based on CPLD Precision Constant Current Source Technology
Author:
Affiliation:

(1.Medical Engineering Department, No.305 Hospital of Chinese People's Liberation Army, Beijing 100017, China; ;2.Respiratory Department, No.305 Hospital of Chinese People's Liberation Army, Beijing 100017, China; ;3.Biomedical Engineering College, Southern Medical University, Guangzhou 510515, China)

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    利用呼吸感应体积描记(RIP)技术测量呼吸参数已得到了越来越广泛的认可,针对人体呼吸运动引起传感线圈的电感量变化非常小(小于几百nH),一般系统难以保证精确性和适用性问题;硬件设计采用了高精度恒流源,锁相放大和自动增益控制放大电路,并详细分析了各部分电路的特性,包括在平静呼吸和用力呼吸模式下分别对14位测试者测量潮气量和呼吸频率,并且与德国耶格公司的肺功能仪所测量的结果对比;实验证明RIP系统在这两种模式下测量潮气量和呼吸频率等呼吸参数与速度描记仪存在显著的正相关关系,说明该系统工作良好、测量精确,操作简单、便于携带和低功耗,未来可以在临床上推广应用于长时间动态的呼吸监测。

    Abstract:

    Respiratory inductive plethysmography(RIP) technology for measuring parameters of breathing has been widely accepted. However, due to small changes (less than few hundred nH) of the inductance caused by the human respiratory movements,general stimulating signal that stimulates the inductive coils can hardly guarantee the precision and the suitability of RIP system. To solve this problem ,this paper present a design combined the technologies of the precision constant current source, the lock-in amplifier and the automatic gain control amplifier. The electric circuit characteristics of the design was also elaborated in this paper. With the breathing measuring experiment on 14 subjects which measured tidal volume and breath rate under quiet breath and forced breath and compared with the result from the pneumotachograph manufactured by JAEGER. The results show that breath rate and other respiratory parameters from our system have a high positive correlation with the those from pneumotachograph.The system was proved to be working at stable and measuring in a better precision.,due to the simple operations, convenience in carrying and the low-power , the system will meet the future demands on the long time ambulatory respiratory monitoring in the clinical. 

    参考文献
    相似文献
    引证文献
引用本文

姬军,黄增跃,沙杭,王丹,刘勇谋.基于CPLD高精度恒流源技术的呼吸体积描记系统的设计计算机测量与控制[J].,2014,22(10):3347-3349.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2015-01-15
  • 出版日期:
文章二维码